Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference, RTSD workshop, Dresden, October 2008 B. J. Heismann a),b) , D. Henseler a) , P. Hackenschmied a) , M. Strassburg a) , D. Niederlöhner a) , S. Janssen a) , S. Wirth a) a) Siemens Healthcare, Forchheim, Germany b) University of Erlangen-Nuremberg, Erlangen, Germany

Transcript of Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution...

Page 1: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging

IEEE Conference, RTSD workshop, Dresden, October 2008 B. J. Heismanna),b), D. Henselera), P. Hackenschmieda), M. Strassburga), D. Niederlöhnera), S. Janssena), S. Wirtha)

a)Siemens Healthcare, Forchheim, Germany b)University of Erlangen-Nuremberg, Erlangen, Germany

Page 2: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 2 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

Medical imaging devices, X- and γ-ray based

Single Photon Emission Computed Tomography

Radiography Computed Tomography

Page 3: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 3 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

Direct conversion detectors and Medical Imaging

SPECT § Dedicated cardiac, scintimammography and small-animal imaging

prototypes exist, improved energy resolution reported

CT § Prototype counting systems developed § Full CT flux rate is found to be a major challenge

CZT material § Defects, e.g. Te inclusions, affect the dynamic electrical field, charge

trapping and resulting spectral behavior § Polarisation occurs at high flux (>107 qt / smm2)

K.B. Parnham, S. Chowdhury, J. Li, D.J. Wagenaar and B.E. Patt, „Second-Generation, Tri-Modality Pre-Clinical Imaging System“, in: Bo Yu, 2007 IEEE Nuclear Science Symposium Conference Record, M06-29 (2006) D .J.Wagenaar, J. Zhang, T. Kazules, T. Vandehei, E. Bolle, S. Chowdhury, K. Parnham, B. E. Patt, “In Vivo Dual-Isotope SPECT Imaging with Improved Energy Resolution”, in: Bo Yu, 2007 IEEE Nuclear Science Symposium Conference Record, MR1-3 (2006) I M. Blevis,M. K. O'Connor, Z. Keidar, A. Pansky, H. Altman and J.W. Hugg, “CZT gamma camera for scintimammography“, Phys. Med, Vol. 21, Supl. 1, p 56-59, (2006) E. Kraft, P. Fischer, M. Karagounis, M. Koch, H. Krueger, I. Peric, N. Wermes, C. Herrmann, A. Nascetti, M. Overdick, and W. Ruetten. Counting and Integrating Readout for Direct Conversion X-ray Imaging: Concept, Realization and First Prototype Measurements. IEEE Transactions on Nuclear Science, 54(2):383–390, (2007) Y. Onishi, T. Nakashima, A. Koike, H. Morii, Y. Neo, H. Mimura and T. Aoki, „Material Discriminated X-Ray CT by Using Conventional Microfocus X-Ray Tube and CdTe Imager“, in: Bo Yu, 2007 IEEE Nuclear Science Symposium Conference Record, M27-2 (2007) D. Moraes, J. Kaplon and E. Nygard, „CERN_DxCTA counting chip“, in: T. Michel, B. Kreisler and G. Anton (Eds), Radiation Imaging Detectors 2007, Proceedings of the 9th International Workshop on Radiation Imaging Detectors (2007) J P Schlomka et al.,“Experimental feasibility of multi-energy photon counting K-edge imaging in pre-clinical computed tomography”, Phys.Med. Biol. 53, pp. 4031–4047, (2008) S. A.Soldner; D.S. Bale and C.Szeles, “Dynamic Lateral Polarization in CdZnTe Under High Flux X-Ray Irradiation”, Nuclear Science, IEEE Transactions on Volume 54, Issue 5, pp.723 – 1727 (2007) E. Bolotnikov, N. Abdul-Jabber, S. Babalola, G. S. Camarda, Y. Cui, A. Hossain, E. Jackson, H. Jackson, J. James, K. T. Kohman, A. Luryi, and R. B. James, “Effects of Te inclusions on the performance of CdZnTe radiation detectors”, in: Bo Yu, 2007 IEEE Nuclear Science Symposium Conference Record, R27-2 (2007) G. S. Camarda, A. E. Bolotnikov, Y. Cui, A. Hossain, S. A. Awadalla, J. Mackenzie, and H. Chen and R. B. James, ”Polarization Studies of CdZnTe Detectors using Synchrotron X-ray Radiation”, in: Bo Yu, 2007 IEEE Nuclear Science Symposium Conference Record, R27-3 (2007) L. Abbene, S. Del Sordo, F. Fauci, G. Gerardi, A. La Manna, G. Raso, A. Cola, E. Perillo, A. Raulo, V. Gostilo and S. Stumbo, “Study of the spectral response of CZT multiple-electrode detectors”, in: Bo Yu, 2007 IEEE Nuclear Science Symposium Conference Record, N24-298 (2007)

Page 4: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 4 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

Motivation

Analyze and compare §  spatial and spectral resolution of pixelized

§ direct conversion semiconductor detectors (e.g. CZT) and §  indirect conversion scintillator detectors (e.g. GdOS)

Establish and verify §  Physical detector models

Use figures of merit §  Modulation transfer function MTF(f) §  Detector response function D(E,E')

Page 5: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 5 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

Indirect conversion detector

CT detector

Page 6: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 6 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

Direct conversion detector

Typical common cathode set-up

Page 7: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 7 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

Figures of merit

Spatial resolution: Modulation Transfer Function

Spectral resolution: § Pulse height spectrum (PHS) § FWHM of emission line § Photo peak fraction

§ Generalization for CT and Rad:

PHS for each input energy: Detector response function D(E,E')

Inte

nsity

E in keV 0 50 100 150 200

CZT

NaI

99mTc spectrum 140 keV

figure taken from: M. N. Wernick and J. N. Aarsvold, ‘Emission Tomography‘, p. 275, Elsevier Academic Press, (2004)

))0(())((

)(PSFFT

xPSFFTfMTF =

FWHM

Page 8: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 8 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

Detector response function

D(i,k)(E, E'):= Macroscopic probability density function for measuring output energy E' at input energy E

This talk:

§  Central pixel D(0,0)(E,E') only

§  Flat-field irradiation

E E‘

Input quantum with energy E, e.g. 130 keV

Detector model

Distribution of output energies

Page 9: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 9 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

Detector modeling

Page 10: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 10 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

Model for indirect conversion detector*

1.  X-ray energy deposition (Monte Carlo look-up table)

2. Light transport probabilities 3. Electronics (noise, linearity, ...)

à Photo-electrons detected in pixel

*Heismann B.J., Pham-Gia K., Metzger W., Niederlöhner D., Wirth S., “Signal transport in Computed Tomography detectors”, Nucl. Inst. and Meth. in Phys. Res., A, 591, pp. 28–33 (2008

P2 P1 P3

IPhoto

channel

Page 11: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 11 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

Model for direct conversion detector (I)

1.  X-ray energy deposition (Monte Carlo look-up table)*

2.  FEM modeling of charge transport**,***

3.  Pulse train signals on electrodes

4.  Electronics (noise, shaping, comparator, ...)

à Pixel count readings

comparator

* J. Giersch and J. Durst, Monte Carlo simulations in X-ray imaging, NIM A, Vol. 591, pp. 300 (2008) ** J. D. Eskin et.al., Signals induced in semiconductor gamma-ray imaging detectors, J. Appl. Phys., Vol.85, pp.647 (1999) *** B. Kreisler et.al., Generalised Adjoint Simulation of Induced Signals in Semiconductor X-Ray Pixel Detectors, submitted to J. of Instrumentation, (2008)

counts

Page 12: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 12 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

Model for direct conversion detector (II)

Configuration for…

… good energy resolution … low spatial resolution

… high spatial resolution … low energy resolution

… good high flux performance

large pixel

long shaping time

short shaping time

counts

counts

small pixel

Parameters for simulation study: CZT with 2mm thickness and 700V bias - small pixel area (pitch = 450µm) - short shaping time (τ = 50ns) - electronic noise equivalent to 3 keV

Page 13: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 13 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

Results

Page 14: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 14 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

Results indirect conversion CT detector

CT detector model reproduces experimental results on

§ Light output and sensitivity profile* § MTF** § DQE

with small deviations in the percent range.

*Heismann B.J., Pham-Gia K., Metzger W., Niederlöhner D., Wirth S., “Signal transport in Computed Tomography detectors”, Nucl. Inst. and Meth. in Phys. Res., A, 591, pp. 28–33 (2008)

**Paper M06-257 : S. Wirth, B.J. Heismann, D. Niederlöhner, L. Bätz, W. Metzger, K. Pham Gia, Simulations and Measurements of the Modulation Transfer Function of Scintillator Arrays, IEEE Conference, Dresden (2008)

Page 15: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 15 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

MTF

Indirect conversion detector § mid-frequency drop due to optical

crosstalk § almost lossless recovery possible

Direct conversion detector § close to ideal sinc behavior § charge sharing affects

spectral behavior

Page 16: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 16 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

E in keV

Central pixel response D(0,0)(E,E')

E in keV

GdOS, indirect conversion CZT, direct conversion

Main branch (nearly) complete energy detection

Side branches partial energy detection

due to fluorescence energy transfer

light tailing

E’ in

opt

. pho

tons

on

PD

E’ i

n ke

V

Page 17: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 17 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

Integrating scintillator detector: Generalized Swank factor

( )( )''

')(

1)(22 EE

ESNRSNR

EEf

in

out

σα +==

SNR decrease by signal transport variance*,**

with α(E): detection efficiency

1000 2000 3000 40000

0.01

0.02

0.03

0.04

D(E

',E)

30 keV

1000 2000 3000 40000

0.005

0.01

0.015

0.02 50 keV

1000 2000 3000 40000

0.005

0.01

0.015

0.02 60 keV

1000 2000 3000 40000

1

2

3

4x 10

-3

E' [Photons on PD]

130 keV

* M. Rabbani, R. Shaw, and R. Van Metter, ‘Detective quantum efficiency of imaging sytems with amplifying and scattering mechanisms,’ J. Opt. Soc. Am. A, vol. 4, pp. 895-901, (1987)

**Heismann B.J., Pham-Gia K., Metzger W., Niederlöhner D., Wirth S., “Signal transport in Computed Tomography detectors”, Nucl. Inst. and Meth. in Phys. Res., A, 591, pp. 28–33, (2008)

E' in light photons on photodiode

D(E=const,E')

drop mainly due to fluorescence

escape and light tailing

f(E)

E in keV

Page 18: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 18 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

E’ in

keV

Eth2

Direct conversion: Energy binning

Eth1 (noise cut-off)

false low energy bin counts

E in keV

Page 19: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 19 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

Dual-energy binning

§ 120kV tungsten spectrum § Eth1 = 15 keV, Eth2 = 55 keV

Fluorescence escape: low energy photon loss if threshold 1 too high

Effective spectra show overlapping region

E in keV

Inte

nsity

Page 20: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 20 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

Conclusion

§  Indirect and direct conversion (e.g. CZT) detector models established §  Figures of merit for comparison: MTF and detector response D(E,E')

Spatial resolution MTF §  Indirect conversion: Mid-frequency drop due to optical crosstalk §  Direct conversion: Close to ideal sinc function

Spectral resolution D(E,E') §  Main linear, fluorescence and differential branches visible §  Indirect conversion scintillator detector: § Additional light tailing and optical cross-talk §  Integrating: Energy-dependent Poisson excess noise

§  Direct conversion semiconductor detector: § Counting yields energy information (full or binned) § Binned: Significant spectral overlap by low energy shift

Page 21: Spatial and spectral resolution of semiconductor detectors in ......Spatial and spectral resolution of semiconductor detectors in medical X-ray and gamma ray imaging IEEE Conference,

Page 21 Oct. 2008 Siemens Healthcare B. J. Heismann et al.

Thank you!

Acknowledgements Jürgen Durst Björn Kreisler Thilo Michel

Michael Balda

Contact